By Lisa Voronkova, CEO of OVA Solutions. Updated on November 25, 2026.
In short
A biocompatible material is one that does what it is meant to do in the body without causing an unacceptable reaction, in a specific device and a specific use. No material is biocompatible on its own: titanium in a bone screw and titanium in a blood pump are judged separately, and so are the same polymer before and after molding and sterilization. Regulators judge the finished device through a biological evaluation under ISO 10993-1, run inside ISO 14971 risk management. The sixth edition, ISO 10993-1:2025, was published on November 18, 2025; FDA recognized it in part on May 25, 2026 and accepts the 2018 edition until July 1, 2029. For a long-term implant in tissue or bone, FDA’s guidance lists ten biological effects to address, and eleven for an implant in contact with blood.
What biocompatible means
The definition biomaterials science has used since the 1980s, associated with David Williams, is the ability of a material to perform with an appropriate host response in a specific application. The last three words carry the meaning. A material that is fine on intact skin for a day may be a problem in bone for ten years; one that is fine in bone may activate clotting in blood.
For long-term implants the practical goal is usually modest. In a review of more than 50 years of implant experience, Williams concluded that the main requirement for a material in long-term contact with tissue is to do no harm, achieved through chemical and biological inertness, and that attempts to add biological activity to such materials had rarely succeeded clinically. Tissue engineering and drug or gene delivery are the exception, where the material is meant to interact with cells.
Biocompatibility is also not the same as safety in general. Wear particles, corrosion, fatigue fracture and heating in an MRI scanner are evaluated through mechanical and other testing, even though they often end in a biological reaction.
Biocompatible materials used in implants
Implant materials are bought to standard specifications that fix chemistry, microstructure and mechanical properties. The ASTM specifications below are the common ones in the US; most are recognized by FDA, sometimes in an older edition than the current one.
| Material | Specification | Typical implant uses | Watch for |
|---|---|---|---|
| Titanium alloy Ti-6Al-4V ELI | ASTM F136 (current F136-26; FDA recognizes F136-13) | Bone plates and screws, spinal rods, joint stems | Notch sensitivity in fatigue; wear if used as a bearing surface |
| Unalloyed titanium, grades 1 to 4 | ASTM F67-24 | Dental implants, plates, housings of active implants | Lower strength than the alloy |
| Cobalt-chromium-molybdenum | ASTM F75-23 (castings), F1537-20 (wrought) | Joint bearing surfaces, femoral components | Metal ions and wear debris at bearing surfaces |
| Stainless steel 316L (implant grade) | ASTM F138-19 | Temporary fracture fixation, wires | Corrosion over long implantation; nickel content |
| PEEK | ASTM F2026-23 | Spinal cages, suture anchors | Radiolucent, so implants carry markers; bonding to bone needs surface treatment |
| UHMWPE | ASTM F648-21 | Bearing liners in hip and knee replacements | Wear particles and oxidation over time |
| Nickel-titanium (nitinol) | ASTM F2063-26 (FDA recognizes F2063-18) | Stents, guidewires, superelastic implants | Nickel release, surface finish, fatigue |
| Additively manufactured Ti-6Al-4V ELI | ASTM F3001-14(2021), F3302-18 for finished parts | Porous orthopedic implants, patient-specific parts | Powder residue in pores, cleaning validation, build-to-build variation |
Ceramics such as alumina and zirconia serve as bearing heads and dental restorations, and silicones and polyurethanes insulate leads and form soft-tissue implants. Each comes in grades meant for long-term implantation and grades that are not; a supplier’s general medical grade often excludes implants over 30 days, so read the supplier’s statement of use.
Why a biocompatible material does not make a biocompatible device
FDA and ISO 10993 evaluate the device in its final, finished form, after every process it goes through. A certificate for the raw bar or resin is an input, not the answer. Between the material and the patient, the device picks up:
- machining fluids, polishing compounds and cleaning residues;
- mold release agents, colorants and additives in molded parts;
- changes from sterilization, such as residues from ethylene oxide or radiation effects on polymers;
- adhesives, coatings and inks, each a material of its own;
- wear and corrosion products formed in use, including galvanic effects where two metals touch;
- loose powder trapped in additively manufactured lattices.
That is why chemical characterization under ISO 10993-18 and a toxicological risk assessment under ISO 10993-17 now sit at the center of most evaluations: they show what can come out of the finished device and whether the amount matters. It is also why a later change of supplier, process or sterilizer can reopen the evaluation, a point to settle in design transfer rather than after launch.
How a biological evaluation works under ISO 10993-1:2025
The 2025 edition is built around risk management. Its title now speaks of evaluating biological safety within a risk management process, and the work runs in this order:
- describe the device, each component that touches the body, the tissue it touches and the exposure duration;
- characterize the materials and the manufacturing, including chemistry where needed;
- identify biological hazards and the biological effects that could follow;
- collect existing data, such as clinical history of the same material in the same use, supplier data, literature and earlier tests;
- test only to fill the gaps, choosing methods from the other parts of ISO 10993;
- assess the biological risk, decide whether it is acceptable and write the biological evaluation report.
Exposure duration is now counted in contact days. Advisera’s example: a device used for 10 minutes a day on 10 days counts as 10 contact days, not 100 minutes, so some devices once treated as limited contact move into prolonged or long-term categories. The duration categories themselves stay at limited (up to 24 hours), prolonged (over 24 hours up to 30 days) and long-term (over 30 days).
Existing data counts. The FDA entry for the 2025 edition notes that the standard does not require retesting of marketed devices with established safety profiles, and a material with a documented history in the same tissue, duration and processing can close several effects without new animal studies.
Which biocompatibility tests an implant needs
FDA’s guidance on ISO 10993-1, issued on September 8, 2023, sorts devices by contact (surface, externally communicating or implant) and duration, and lists the effects each must address by data, testing or a written rationale. For long-term implants its table gives:
| Biological effect | Implant in tissue or bone, over 30 days | Implant in contact with blood, over 30 days |
|---|---|---|
| Cytotoxicity | Yes | Yes |
| Sensitization | Yes | Yes |
| Irritation or intracutaneous reactivity | Yes | Yes |
| Acute systemic toxicity | Yes | Yes |
| Material-mediated pyrogenicity | Yes | Yes |
| Subacute and subchronic toxicity | Yes | Yes |
| Genotoxicity | Yes | Yes |
| Implantation | Yes | Yes |
| Chronic toxicity | Yes | Yes |
| Carcinogenicity | Yes | Yes |
| Hemocompatibility | No | Yes |
“Address” does not always mean “test”. Carcinogenicity and chronic toxicity are often answered with chemistry, toxicology and material history rather than a lifetime animal study. Two details from the guidance save rework: a device used for more than 14 days should not be assessed only with a 14-day test, and material-mediated pyrogenicity needs its own study rather than a reading taken from another test.
The ISO 10993 parts you will meet
| Part | What it covers | Current edition | FDA recognition |
|---|---|---|---|
| 10993-1 | Evaluation and testing within a risk management process | 2025 | 2-313, partial; 2018 edition 2-258 until July 1, 2029 |
| 10993-3 | Genotoxicity, carcinogenicity, reproductive and developmental toxicity | 2026 | 2014 edition, 2-228, partial |
| 10993-4 | Interactions with blood | 2017 with amendment 1 of 2025 | 2-311, complete |
| 10993-5 | In vitro cytotoxicity | 2009, confirmed 2022 | 2-245, complete |
| 10993-6 | Local effects after implantation | 2026 | 2016 edition, 2-247, complete |
| 10993-10 | Skin sensitization | 2021 | 2-296, partial |
| 10993-11 | Systemic toxicity | 2017 | 2-255, complete |
| 10993-12 | Sample preparation and reference materials | 2021 with amendment 1 of 2025 | 2-314, complete |
| 10993-17 | Toxicological risk assessment of device constituents | 2023 with amendment 1 of 2025 | 2-303, partial |
| 10993-18 | Chemical characterization of materials | 2020 with amendment 1 of 2022 | 2-298, partial |
| 10993-23 | Irritation | 2021 with amendment 1 of 2025 | 2-291, partial |
What changed in 2025 and 2026
The sixth edition of ISO 10993-1 reorganized the whole document. According to its foreword, it adds guidance on calculating exposure duration, on describing the device and on identifying biological hazards; renames biological endpoints as biological effects and revises their list; replaces the term externally communicating with a description of the tissue each component touches; renames implantation effects as local effects after tissue contact, since some devices that are not implants need that evaluation; and moves most of the old Annex A into the main text.
FDA recognized the 2025 edition in part with an entry date of May 25, 2026, published in recognition list 066. It did not recognize clause 6.9 on biological risk estimation, pointing to ISO 14971:2019 instead, nor the phrase about consumer products in clause 6.5.11.3, and it noted that the extra genotoxicity expectations in the 2025 tables may not match its own table for every prolonged-contact device. Declarations to the 2018 edition remain acceptable until July 1, 2029, so a program that starts now can choose, but a submission after mid-2029 cannot.
- ISO 10993-3:2026 and ISO 10993-6:2026 are published, while FDA still recognizes the 2014 and 2016 editions.
- FDA accepts declarations to the 2002 edition of ISO 10993-17 only until December 20, 2026; after that, toxicological risk assessments follow the 2023 edition.
- FDA’s draft guidance on chemical analysis for biocompatibility, published in September 2024, had no final version in the Federal Register by November 25, 2026.
How long biocompatibility testing takes, and where to run it
Turnaround depends on the lab and the test. One accredited GLP lab, GBA Medical Device Services, quotes these times from sample receipt, with about a week more for GLP documentation:
| Test | Typical turnaround |
|---|---|
| Cytotoxicity, MTT or MEM elution | 4 to 5 weeks |
| Extractables and leachables chemistry | 4 to 8 weeks |
| Irritation or intracutaneous reactivity | 6 to 7 weeks |
| Acute systemic toxicity; rabbit pyrogen test | 6 to 7 weeks each |
| Sensitization, guinea pig maximization | 8 to 9 weeks |
| Sensitization, local lymph node assay | 12 to 13 weeks |
| Genotoxicity, Ames test | 3 to 4 months |
| Genotoxicity, in vitro chromosomal aberration or micronucleus | 6 to 7 months |
Many tests can run in parallel, so the critical path is usually the longest study plus the time to build test articles in their final, sterilized form. Under FDA’s ASCA program, labs accredited for listed biocompatibility methods report results FDA accepts without re-review unless it finds a reason not to. The listed methods cover parts 4, 5, 10, 11, 12 and 23 of ISO 10993 and the USP rabbit pyrogen test; implantation under ISO 10993-6 is not in ASCA.
Active and smart implants: electronics inside the body
An implant that senses, stimulates or communicates adds a second problem: the electronics must never touch tissue, and the package around them must stay sealed for the life of the device. The usual answer is a hermetic titanium or ceramic housing with feedthroughs for the electrodes, or for smaller parts an encapsulation whose own biocompatibility and moisture barrier are tested as part of the device. Every material on the outside, from the header polymer to the lead insulation and the electrode metal, joins the biological evaluation.
Service life shapes the architecture. On a wearable controller for a neural implant we built, the implant stays in the body while the hub on the skin can be charged, replaced and updated, so every function that might change belongs in the hub. Skin-contact parts have their own biocompatibility questions; our article on wearable ECG and PPG patches covers adhesives and electrodes worn for days. In orthopedics we test mechanics the same way we test electronics, as in the durability validation of a cranial fixation driver with 100 robotic insertions under constant torque.
Mistakes that delay biocompatibility
| Mistake | What it costs | What to do instead |
|---|---|---|
| Testing prototype parts | Results that do not represent the device; tests repeated on production parts | Test articles made by the final process, cleaned, packaged and sterilized |
| Relying on the supplier’s certificate | Gaps found by the reviewer, months late | Use supplier data as one input and evaluate the finished device |
| Running every test in the table | Animal studies and months that existing data could have avoided | A biological evaluation plan that starts with chemistry and existing data |
| Changing a material or process after testing | A new evaluation, sometimes new tests | Lock materials, suppliers and sterilization before test articles are built |
| Counting minutes instead of contact days | A device placed in the wrong duration category | Count contact days as the 2025 edition describes |
| Ignoring the transition dates | A submission built on an edition FDA no longer accepts | Check the recognition entry for every standard in the plan |
What OVA does here, and where our responsibility ends
We design the electronics, firmware and mechanics of medical devices, including orthopedic devices and the wearable parts of implant systems, under an ISO 13485:2016 certified quality system (certificate details). We choose materials with the biological evaluation in mind, document materials and processes for it and build test articles in their final form. We are not a test lab and not a regulatory consultancy: the tests run at accredited labs, and the biological evaluation report and submission stay with you or your regulatory consultant.
We agree a fixed price for the product before development starts. The first paid step is the Device-to-Market Blueprint: 4 to 6 weeks, $25,000 fixed, credited in full to development if you sign within 90 days.
This page is engineering guidance, not a regulatory or toxicological opinion.
Questions
What is a biocompatible material?
A material that performs its function in the body without an unacceptable biological reaction, in a specific device and use. Biocompatibility belongs to the finished device in its application, not to a material on its own.
Which materials are biocompatible for implants?
The common ones are titanium and Ti-6Al-4V ELI, cobalt-chromium-molybdenum alloys, implant-grade 316L stainless steel, nitinol, PEEK, UHMWPE, alumina and zirconia ceramics and implant-grade silicones, each bought to a standard specification. The finished device still needs its own evaluation.
Is titanium biocompatible?
Titanium and its ELI alloy have a long record in bone and soft tissue, which is why they dominate orthopedic and dental implants and the housings of active implants. A titanium device is still evaluated as made, with its surface, residues and sterilization.
What is ISO 10993?
A series of standards for the biological evaluation of medical devices. Part 1 sets the framework within risk management; the other parts describe tests and chemistry, from cytotoxicity (part 5) to implantation (part 6) and chemical characterization (part 18).
Which biocompatibility tests does an implant need?
For an implant in tissue or bone for more than 30 days, FDA’s guidance lists ten effects to address, from cytotoxicity to carcinogenicity, and adds hemocompatibility for blood contact. Many are answered with chemistry and existing data rather than new animal tests.
How long does biocompatibility testing take?
Single tests take from about 4 weeks for cytotoxicity to 6 or 7 months for some genotoxicity studies, and implantation and chronic studies longer. Running tests in parallel on final, sterilized articles keeps the program to a few months.
Does FDA accept ISO 10993-1:2025?
In part, since May 25, 2026 (recognition number 2-313). Clause 6.9 on biological risk estimation is not recognized. FDA also accepts the 2018 edition until July 1, 2029.
Related pages
Sources
- ISO 10993-1:2025, biological evaluation of medical devices, part 1 (iso.org)
- FDA recognized consensus standards: ISO 10993-1:2025, recognition number 2-313
- FDA recognized consensus standards: ISO 10993-1:2018, recognition number 2-258
- FDA recognized consensus standards: ISO 10993-17:2023, recognition number 2-303
- FDA recognized consensus standards database
- Federal Register, recognition list 066 (FR document 2026-17229)
- Federal Register, FDA guidance on the use of ISO 10993-1, September 8, 2023
- FDA, Use of International Standard ISO 10993-1 (guidance document)
- Federal Register, draft guidance on chemical analysis for biocompatibility assessment, September 20, 2024
- Federal Register, ASCA program notice, September 23, 2024
- Advisera, ISO 10993-1:2025, what has changed
- GBA Medical Device Services, biocompatibility testing flyer 2025
- Williams DF, On the mechanisms of biocompatibility, Biomaterials 2008;29(20):2941-53
- ASTM F136, F67, F75, F1537, F138, F2026, F648, F2063, F3001, F3302 (ASTM store)