By Lisa Voronkova, CEO of OVA Solutions. Updated on November 24, 2026.
In short
A wearable ECG or PPG patch rarely fails on the bench. It fails between prototype and product, when the signal changes with movement, the electrode or adhesive gives out before the battery does, Bluetooth drops data, the optical reading shifts with skin and placement, and the claims run ahead of the evidence. FDA treats most ECG patches as Class II devices cleared through 510(k), under product codes such as DSH or DXH; a pulse oximetry function adds the oximeter code DQA. Below are four OVA wearable programs, what each one broke and fixed, and the FDA records, standards and tests that matter in 2026.
Four OVA programs and what each one broke
About 30% of our ongoing projects in mid-2026 were wearables (wearable medical device development). The first three below were worn on the body. The fourth is a wireless EEG acquisition board, included because its analog front end faces the same noise problem as an ECG patch. Client names are not disclosed; every number comes from the published case study or engineering note linked in each section.
1. A wrist monitor with ECG, SpO2 and a real blood pressure cuff
The brief was a home remote patient monitor that measures blood pressure with an inflatable wrist cuff, plus ECG, oxygen saturation, pulse and skin temperature, and exchanges data with a server that can configure the device remotely. A team of 14 took it from requirements through production, device assembly and quality control, on an i.MX8M with Yocto Linux, dual-band Wi-Fi and Bluetooth 5.0.
On paper it was five independent subsystems sharing a battery. In the enclosure they degraded each other. The cuff pump is a motor on the same chassis as an ECG front end that resolves microvolts and photodiodes that read the pulse wave, so ECG and SpO2 could not be sampled during inflation. The processor with Wi-Fi active heated the small volume next to the temperature sensor, and the 5 GHz radio conflicted with the ECG front end.
Most of the fix was a schedule, not a filter. We built an interference matrix of every sensor pair before the block diagram, then turned concurrency into a sequence: no ECG or SpO2 during inflation, radio transmissions outside the ECG capture window, and a settling delay before each temperature reading, with shielding and layout discipline for the radio. A thermal break between the compute stack and the sensor pod, with the temperature sensor on its own skin-contact island, took care of the heat. In our tests blood pressure and pulse stayed within 5% of reference devices.
The lesson for an ECG or PPG patch: the measurement schedule sets the duty cycle, the duty cycle sets the power budget, the power budget sets the cell, and the cell sets the size. Fix the schedule during architecture. Case study and engineering note.
2. A single-use glucose patch with a two-week wear target
A pharma company asked us for a proprietary continuous glucose monitoring patch: a skin-worn, single-use electrochemical sensor with its own needle and chemistry, an applicator, a two-week wear target and Bluetooth. We delivered it in eight phases to a clinical-ready prototype, 11,600 engineering hours over one and a half years. The device is now in clinical trials.
Skin got a prototype round of its own. The first round built a batch of 100 units for bench and accuracy work. The second round covered optimization and the applicator, plus tissue and model testing and small trials to find a design that does not irritate the skin during long wear. The clinical phase took about 2,000 hours on its own: about 30 devices, the study software and protocol, support during the study and analysis of the results, followed by a fourth prototype updated on what the study found.
The lesson: wear time is a design input, not a test result. Plan a prototype round for skin, as this program did, run it with production materials, and budget the clinical phase as engineering work. Case study.
3. A stimulator whose disposable electrode decided everything
For a portable, rechargeable, Bluetooth-controlled facial muscle stimulator we designed the electronics and firmware; the client’s device received FDA clearance. The module is sold once. The disposable pre-gelled electrode, attached magnetically, is bought for as long as therapy lasts, and it set the main constraint on our electronics even though it was outside our scope.
A fresh pad and the same pad after eight applications present very different impedance. A constant-voltage output stage lets the dose fade as the gel dries; a constant-current stage holds the dose but raises compliance voltage and concentrates current in whatever contact area remains. The rule we wrote down from this program: specify the output stage against the electrode’s end-of-life impedance, and measure impedance in the device every session so it can derate, warn or refuse, and log the reason.
The lesson for ECG electrodes and PPG optics: specify the front end against an aged interface, measure contact quality in the device, and qualify a second source of the skin-contact material before submission, because a late change can reopen human factors and the clearance rationale. Case study and engineering note.
4. An EEG board that reads microvolts next to a radio
For a client’s wireless EEG monitoring device we designed the acquisition board: the analog front end, the digital and radio sections, and a board ready for serial production. It took 14 months from January 2022 and about 800 development hours, 40% below the initial estimate, and the client’s device later received FDA marketing authorization.
EEG signals are tens of microvolts, far smaller than a surface ECG, and digital circuits and the radio on the same board inject noise that can bury them. The usual answer is more filtering stages. We had analog and digital engineers working side by side from day one and replaced extra filters with a ground-plane layout tested against the noise, which cut BOM cost by 15% and improved signal quality; manufacturing cost came in 30% below target.
The lesson: on a biopotential front end, layout and grounding beat added parts, and integration testing with crude prototypes beats waiting for perfect subsystems. Case study.
Front end, radio and battery: decisions to make before the first board
Three choices set the noise floor, the power budget and the size of a patch, and all three are cheaper to make on paper than on a second board run. Make them from the claim you will put in front of FDA, not from the evaluation board that happened to be on someone’s desk. Whether one ECG lead is enough is a clinical question, so answer it with your clinical advisor first.
| Analog front end | What it gives you | When it fits |
|---|---|---|
| TI ADS1292R | Two 24-bit ECG channels at about 335 µW per channel, with respiration measured by impedance | Two leads, or ECG plus breathing rate from the same electrodes |
| ADI MAX30001 | One ECG channel with bioimpedance; ADI lists the ECG channel at about 85 µW from a 1.1 V supply | Single-lead patches where every microwatt turns into battery size, and battery size into patch size |
| ADI MAX86176 | Two PPG channels (up to 6 LEDs and 4 photodiodes) and a single-lead ECG, sampled in sync | Algorithms that need ECG and PPG on one clock, for example pulse transit time |
The chip is half of the noise story; the board is the other half. We revised an existing electrocardiograph built on an STM32 with an ADS1298 front end and defibrillation and ESD protection, while the client’s own software had to stay untouched, and on another EEG board we optimized the work was multilayer routing for low noise. On a patch the radio and the electrode inputs sit a few millimeters apart, so the same discipline applies from the first schematic.
For the radio, a new design can start on the Nordic nRF54L15, the successor to the nRF52 family: a 128 MHz Arm Cortex-M33 with a RISC-V coprocessor, 1.5 MB of non-volatile memory and 256 KB of RAM. The memory matters on a patch that has to keep recording while the phone is out of range. We used it on a wearable hub for a medical implant, a controller that turns gait data into stimulation patterns, running Zephyr RTOS with Bluetooth 5.4. The smart crutch we built for a client, which measures how much load a patient puts through it, runs on an nRF52. If you already have validated nRF52 firmware, moving costs real money; if you start from zero, start on the newer part.
For the battery, start from the wear time and work backwards. The iRhythm Zio monitor, cleared on August 15, 2025 (K243650), records continuously for up to 14 days, and buyers will compare a new patch to that number. When we were asked to quote a disposable chest patch with ECG, cuffless blood pressure, SpO2, temperature, respiration, an accelerometer, cellular and Bluetooth and 14 to 15 days of wear, our CTO’s rule was no quote before a proof of concept on the cellular chip and the battery, about 500 hours over 2 to 2.5 months. If the modem eats most of a two-week budget, every other line of the estimate is fiction. Lithium cells in a body-worn device are normally expected to meet IEC 62133-2 (2017, amended 2021), so pick a cell whose supplier already has that test report.
Software is where late surprises get expensive. We once wrote functional firmware for an AI stethoscope; the client then hired a separate US company that spent about two years reworking it into compliance-ready software, risk management included. Firmware that runs and firmware that is ready for a submission are two different deliverables: agree which one you are buying on the day you sign.
What FDA expects from an ECG or PPG patch in 2026
A patch is not one regulatory category. Recording ECG, analyzing rhythm, transmitting signals, measuring pulse rate and measuring SpO2 can each change the classification and the evidence. The codes below are the ones recent patch submissions used; all are Class II and normally reach the market through 510(k) clearance. The intended use and the claims decide which apply, not the sensor. Our 510(k) clearance time by product code tool shows how long FDA took for each.
| Product code | FDA name and regulation | Where it fits on a patch |
|---|---|---|
| DSH | Recorder, magnetic tape, medical; 21 CFR 870.2800 | Used by modern ambulatory ECG recorders despite the old name |
| MWJ | Electrocardiograph, ambulatory, without analysis; 21 CFR 870.2800 | ECG recording with no automated analysis, for example the wearable ECG recorder K254119, September 2026 |
| MLO | Electrocardiograph, ambulatory, with analysis algorithm; 21 CFR 870.2800 | ECG recording plus a rhythm analysis algorithm |
| DXH | Transmitters and receivers, electrocardiograph, telephone; 21 CFR 870.2920 | ECG transmission; no telephone line needed despite the name |
| DQK | Computer, diagnostic, programmable; 21 CFR 870.1425 | Analysis software that reads the recorded ECG |
| DQA | Oximeter; 21 CFR 870.2700 | A medical SpO2 function, not every optical pulse reading |
| DRG | Transmitters and receivers, physiological signal, radiofrequency; 21 CFR 870.2910 | Multi-parameter patches that transmit several signals |
Three ECG patches FDA cleared recently, and what their clearances actually cover
| Device | 510(k) and decision date | What the clearance covers |
|---|---|---|
| Rhythm Master ECG Patch (Smwmed) | K252389, April 23, 2026, DSH | Continuous ECG recording for 3 or 7 days depending on the model; reusable recorder, single-use electrodes; no real-time monitoring or automated interpretation |
| MEMO Patch M (Huinno) | K243438, December 5, 2025, DXH | Single-channel ECG recorded, stored and transferred for later review, up to 8 days; electrodes replaced every 24 hours or when adhesion fails; no automated analysis |
| AT-Patch ATP-C130 / ATP-C70 (ATsens) | K242583, May 2, 2025, DSH (with DQK and MLO) | Up to 14 or 7 days of continuous ECG, analyzed after recording and reviewed by a clinician; Bluetooth used to check signal quality, not for real-time monitoring; nonsterile |
Three things in these records change how a patch should be designed. Recording time and electrode time are different numbers: a patch can record for eight days while its electrodes are changed daily. A Bluetooth radio does not make a device a real-time monitor: none of these three clearances covers real-time monitoring. And a skin patch does not have to be sterile. A fourth record shows where PPG fits: the Perin Health System, cleared on January 22, 2026 under the multi-parameter code DRG with the oximeter code among its others, schedules PPG and SpO2 measurements on a chest patch.
For the optical side, FDA published a draft guidance on pulse oximeters for medical purposes in January 2025. It proposes evaluating performance across skin pigmentation and covers reflectance sensors as well as fingertip clips. It is still a draft, so treat it as the direction of travel: validate the final sensor, placement and algorithm across skin tones, and do not carry fingertip results over to a chest or wrist. FDA lists it as a draft on its guidance page.
Wellness product or medical device: where the line sits in 2026
FDA revised its general wellness guidance on January 6, 2026 (summary by Ropes & Gray). Non-invasive wearables that estimate blood pressure, SpO2 or glucose may now be wellness products if they make no disease claims, use no clinical thresholds and send no disease-management alerts; they may show values and trends and suggest seeing a clinician. WHOOP is the clearest recent example: FDA sent a warning letter in July 2025 about its Blood Pressure Insights feature, WHOOP dropped wording that implied clinical accuracy, replaced color-coded cutoffs with a continuous gradient and labeled the feature for wellness and trend tracking, and FDA issued a closeout letter on June 17, 2026.
The wellness route is faster, and it also decides the business. A wellness wearable is usually paid for by the consumer; a medical device can be paid for by an insurer. Choose the line on purpose, because the claims, the evidence and the price all follow from it.
Standards to plan for
Which standards apply depends on the functions, the use environment, the power source and the skin contact. Put editions and FDA recognition status in the project’s standards matrix early; FDA’s recognized standards database shows what is recognized and where recognition is partial. Since February 2, 2026, FDA’s Quality Management System Regulation incorporates ISO 13485:2016 by reference, so the design and development file should trace the front end choice, the battery tests against the wear time, the EMC and biocompatibility results and the software documentation back to requirements.
| Standard | What it covers on a patch | When it applies |
|---|---|---|
| IEC 60601-1 | Basic safety and essential performance of medical electrical equipment | Every powered patch |
| IEC 60601-1-2 | Electromagnetic emissions and immunity | Every powered patch; plan essential performance under interference |
| IEC 60601-1-11 | Use in the home healthcare environment | Patches worn at home, even if a clinician applies them |
| IEC 60601-2-47 | Ambulatory electrocardiographic systems | ECG recording patches |
| ISO 80601-2-61 | Pulse oximeter equipment; the third edition came out in April 2026 and the 2017 edition is withdrawn | Medical SpO2 claims, not heart rate alone |
| ISO 10993-1, -5, -10, -23 | Biological evaluation; cytotoxicity, sensitization, irritation (ISO 10993-23, 2021, amended 2025) | Every skin-contacting material; a 7 or 14 day patch is prolonged contact |
| IEC 62304 | Software life cycle; the current edition is 2006 with Amendment 1 (2015), and edition 2 is still a draft | Firmware, the app and analysis software |
| IEC 62366-1 | Usability engineering | Applying, pairing, replacing and removing the patch |
| IEC 62133-2 | Rechargeable lithium cells and batteries | Rechargeable designs |
| ANSI/AAMI EC12 and EC57 | Disposable ECG electrodes; testing rhythm analysis algorithms | Single-use electrodes; analysis claims |
| ANSI C63.27, AAMI TIR69 | Wireless coexistence | Bluetooth and Wi-Fi patches |
Where patches break between prototype and product
A prototype proves that a clean signal is possible on a resting engineer. A product has to deliver usable data from every user, for the whole claimed wear time, after shipping, sweat and a dozen phone reconnects. These are the gaps we look for at the architecture stage, starting with the interference matrix; they are engineering risks, not a ranked statistic.
| Risk | Why the prototype still passes | Evidence to have before design freeze |
|---|---|---|
| Motion artifact | Bench and resting tests hide changes at the skin interface | Usable-data fraction and dropouts during walking, sleep and daily tasks across the intended users |
| Electrode and adhesive wear | A fresh electrode on day one looks perfect | Impedance, noise, detachment and skin response across the full claimed wear time, on aged electrodes |
| PPG and skin | One placement, one skin tone, good perfusion | Accuracy of the final sensor and algorithm across skin pigmentation, placement and low perfusion |
| Power against acquisition and radio | Datasheet currents and a short demo | Measured energy per mode, with reconnects and retransmissions; check it in our battery life estimator |
| Bluetooth data loss | The phone sits next to the patch on the bench | Defined acceptable gaps and delay; tested buffering, timestamps, reconnects and recovery |
| Sweat and water | An IP rating on the enclosure | Adhesion and signal after realistic sweat and aging; IP rating is not proof of wear |
| Reuse, cleaning or sterility | One unit, used once | An early choice of reusable core or single-use patch, with cleaning or sterility validated for that choice |
| Claims | The demo shows a waveform | Claims matched to the evidence: recording, analysis, pulse rate or SpO2, and real time only if tested as such |
A patch with Bluetooth and an app is likely a cyber device under section 524B; the cyber device check shows whether it applies to yours.
What breaks in production, after the prototype works
A prototype is built and worn by engineers. A product is built by a factory and charged by a tired patient with whatever cable is in the kitchen drawer. Two recent Class I recalls, FDA’s most serious category, had nothing to do with sensing:
- In May 2025 Dexcom recalled CGM receivers because the speaker could lose contact with the circuit board, related to the foam or the assembly, so a receiver could fail to sound an alarm. The recall covered more than 700,000 units, with 56 severe adverse events reported (MedTech Dive).
- In November 2025 certain lots of Abbott’s FreeStyle Libre 3 and 3 Plus sensors were recalled for falsely low glucose readings, traced to one production line; FDA’s update linked the issue to 860 serious injuries and 7 deaths (Cardiovascular Business).
Both are assembly and production stories, which makes assembly a design decision. A measurement device is only as good as its worst production unit, so the production test has to measure the thing the device claims to measure, against a reference, on every unit or every lot, not only check that it powers on. Sealing gets the same treatment: for an offender-monitoring security bracelet, OVA paid about $10,000 for a one-day hackathon in which people tried to remove it without triggering the alarm, and nobody succeeded. Sweat is a slower opponent than a motivated person with tools, but it has more time.
On the wrist monitor above we handled production, device assembly and quality control ourselves. Before a patch goes to production, this is the list we would want checked:
- a measured power budget at the real wear time and real radio use
- adhesive and electrode tests on people who are not your engineers
- ingress tests on the sealed housing, charging contacts included
- EMC pre-testing for the home healthcare environment (IEC 60601-1-2)
- an assembly check for every part held in place by pressure, foam or clips
- a production fixture that measures each unit against a reference
- traceability from each unit back to its line, lot and components
Hours and timeline
Wearables vary more than most devices, so take these as reference points from our own programs and our cost calculator, not as a quote.
| Program | Scope | Engineering hours | Duration |
|---|---|---|---|
| EEG acquisition board | Analog front end, digital and radio sections, board ready for serial production | about 800 | 14 months |
| Single-use glucose patch | Discovery to a clinical-ready prototype, including clinical study support | 11,600 | about 1.5 years |
| Class II device that reads body signals, runs on a battery and uses Bluetooth, idea to production-ready design | 3 to 5 prototypes (OVA cost calculator) | 4,340 to 6,040 | 6 to 12 months |
Questions to ask any engineering firm before it builds your patch
These are the questions we would ask in your place, and the ones we expect to answer about our own work. More in how to choose a medical device development firm.
- Which patch or wearable like ours did you take past the prototype stage, and which parts of it did your team own: electronics, firmware, mechanics, algorithms, the app?
- How will you show usable data across the full wear time, and what counts as usable?
- Which FDA product codes and predicates do you see for our claims, and what evidence gap does that leave?
- Does your ISO 13485 certificate cover this work, and who holds design control responsibility: you or us?
- Who owns the firmware, the raw data, the algorithms, the tooling and the design files if we part ways?
- What happens to the data when the phone is out of range, and how is that tested?
- How will PPG accuracy be validated across skin tones on the final hardware?
- What does transfer to a contract manufacturer include: test strategy, fixtures, electrode supply, a pilot build?
What OVA does here, and where our responsibility ends
We design the electronics, firmware and mechanics of wearable medical devices and take them from concept to design for manufacturing, under an ISO 13485:2016 certified quality system (certificate details). We are not a regulatory consultancy and not a contract manufacturer: the regulatory strategy and the submission stay with you or your regulatory consultant, and production stays with your manufacturer.
We agree a fixed price for the product before development starts. The first paid step is the Device-to-Market Blueprint: 4 to 6 weeks, $25,000 fixed, credited in full to development if you sign within 90 days. It answers who uses the patch, who buys it and who pays for it, maps the FDA path and the unit economics, and tests the riskiest subsystem, on a patch usually the front end on real skin or the battery against the wear time.
This page is engineering guidance, not a regulatory opinion. Confirm the product code and pathway with a regulatory specialist or through a Pre-Submission (Q-Sub) with FDA.
Questions
What FDA class is a wearable ECG patch?
Most are Class II devices that reach the market through 510(k) clearance. Recent examples used the product codes DSH (ambulatory recorders), DXH (ECG transmission) and, where rhythm analysis is included, MLO and DQK. The intended use and the claims decide the code, so a patch that adds analysis or real-time monitoring can need different evidence than one that only records.
Does a PPG heart-rate patch need FDA clearance?
It depends on the claims. A low-risk product that only supports general wellness claims may fall under FDA’s general wellness policy, revised on January 6, 2026: a non-invasive wearable may show SpO2 or blood pressure estimates and trends if it makes no disease claims, uses no clinical thresholds and sends no disease-management alerts. Once the patch claims to measure SpO2 for medical purposes, or to detect, monitor or diagnose a condition, it is a medical device; a medical SpO2 function falls under the oximeter code DQA.
How long should an ECG patch record?
The patches FDA cleared from May 2025 to April 2026 that we reviewed record for 3 to 14 days depending on the model. Recording time and electrode time can differ: one cleared patch records for up to 8 days with electrodes replaced every 24 hours.
Why does my patch work on the bench and fail on patients?
Because the bench hides the skin interface. Motion, sweat, aging electrodes, skin tone and placement change the signal, and a phone that moves away from the patch drops data. Test usable-data fraction across the full wear time on the intended users, with production electrodes and adhesives, before the design freezes.
How long does it take to develop a wearable ECG or PPG patch?
Our cost calculator puts a Class II device that reads body signals, runs on a battery and uses Bluetooth, with 3 to 5 prototypes, at 4,340 to 6,040 engineering hours and about 6 to 12 months from idea to a production-ready design with a dedicated team. FDA review, any clinical study and volume production come on top. Our single-use glucose patch took 11,600 hours over one and a half years to a clinical-ready prototype.
Does a skin patch have to be sterile?
Not necessarily. One of the patches FDA cleared in 2025 is not sterile, according to its 510(k) summary, and two of the three we reviewed use a reusable recorder with single-use electrodes. Decide reusable core or single-use patch early, then validate cleaning, reuse or sterility for that choice.
Related pages
Sources
- FDA 510(k) K252389, Rhythm Master ECG Patch, decision April 23, 2026
- FDA 510(k) K243438, MEMO Patch M, decision December 5, 2025
- FDA 510(k) K242583, AT-Patch ATP-C130 / ATP-C70, decision May 2, 2025
- FDA 510(k) K252984, Perin Health System, decision January 22, 2026
- 21 CFR 870.2800, medical magnetic tape recorder (ambulatory ECG)
- 21 CFR 870.2700, oximeter
- FDA, Pulse Oximeters for Medical Purposes, draft guidance, January 2025
- FDA guidance page: Pulse Oximeters for Medical Purposes (draft)
- FDA Recognized Consensus Standards database
- FDA 510(k) K243650, Zio monitor (iRhythm Technologies), decision August 15, 2025
- Texas Instruments, ADS1292R product page
- Analog Devices, MAX30001 product page
- Analog Devices, MAX86176 product page
- Nordic Semiconductor, nRF54L15 product page
- IEC 62133-2, safety of lithium cells and batteries
- ISO 10993-23, tests for irritation
- ISO 80601-2-61, pulse oximeter equipment
- FDA, Quality Management System Regulation (QMSR)
- MedTech Dive, Dexcom recall of CGM receivers that may not sound an alarm
- Cardiovascular Business, FDA update on CGM sensor recalls linked to 860 serious injuries
- Ropes & Gray, FDA updated guidances on general wellness products, January 2026
- FDA, WHOOP Inc. closeout letter, June 17, 2026
- Lisa Voronkova on LinkedIn, ECG and PPG patch electronics under ISO 13485: design decisions, November 9, 2026
- Lisa Voronkova on LinkedIn, Wearable medical devices: what breaks between prototype and product, November 20, 2026
- OVA Solutions case study: wrist remote patient monitor with cuff blood pressure
- OVA Solutions case study: single-use glucose monitoring patch
- OVA Solutions case study: wearable muscle stimulator
- OVA Solutions case study: EEG acquisition board
- Lisa Voronkova, Five vitals on one wrist: the interference problem in multi-parameter wearables
- Lisa Voronkova, The consumable is the product: electronics for an FDA-cleared wearable stimulator